Calibration of a pressure transmitter
Calibration with analog apparatus
Equipment
Pressure Transmitter range (0 to 200 psig )
Pressure source range (0 to 250 or more than)
Multi meter
24vdc power supply
Procedure
Set range of transmitter to required range (0 to 200 psig)
Connect multi meter in series of power supply to the transmitter
Make graduation of the range of transmitter as follows
(for 0 to 200 psig range)
Applied pressure | Multi meter output | %age |
0 psig | 4 ma | 0% |
50 psig | 8 ma | 25% |
100 psig | 12 ma | 50% |
150 psig | 16 ma | 75%0 |
200 psig | 20 ma | 100% |
First adjust the zero (with zero screw) that multi meter must read 4ma
If multi meter is reading less then 4 ma then increase the out put by clockwise moving to the zero set screw
And if multi meter is reading more than 4 ma then decrease the out put by moving the anticlockwise to zero set screw
After adjusting the zero apply 50 psig pressure (as u made graduation) to the transmitter where multimeter must show 8ma if not then adjust it from span screw
After adjusting check
100 psig = 12ma.
150 psig = 16ma
200 psig = 20ma respectively
after adjusting all the readings check two or more time to all readings if no changes occur then your transmitter is calibrated
Calibration with HART communicator (Highway addressable remote transducer)
Equipment
Pressure Transmitter range (0 to 300 psig ) “must be Hart enabled”
Pressure source range (0 to 350 or more than)
Hart communicator (275/375)
24vdc power supply
Procedure
Connect the Hart communicator with the transmitter
Connect its terminals across the 250 Ohms resistor connected in series of power supply to the transmitter
When Hart communicator detects your device then go into its setup
Adjust the range through the option lower/upper range value
Set lower range value to zero (0) and upper range value to 300 psig
Set the zero by zero trim option (given pressure must be zero)
Apply 300 psig pressure to the transmitter and set it the 100% range value of transmitter by span trim option
After this setting check the 0 and 100% out put by connecting also multi meter in the series of the power supply
At 0 psig pressure multi meter must read 4ma
Apply the 300 psig where multi meter must read 20ma
Repeat this practice two or three times your transmitter will be calibrated when it matches its output reading match to the input